National Instruments Corporation
81
NI 4065 Calibration Procedure
23. Call the niDMM Cal Adjust Gain VI.
24. Output –1 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
25. Call the niDMM Cal Adjust Gain VI.
26. Output 0 A on the calibrator.
27. Wait 5 seconds before making a measurement to allow thermal EMF
to settle.
28. Call the niDMM Cal Adjust Offset VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
1
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
1
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: –
1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA