National Instruments Corporation
89
NI 4065 Calibration Procedure
AC Current
100 mA
100 m
99.945 mA
100.055 mA
99.986 mA
100.014 mA
–1 A
1
–1.0011 A
–0.9989 A
–1.00056 A
–0.99944 A
0 A
1
–100
A
100
A
–60
A
60
A
1 A
1
0.9989 A
1.0011 A
0.99944 A
1.00056 A
–2.2 A
3
–2.204241 A
–2.195759 A
–2.203801 A
–2.196199 A
0 A
3
–600
A
600
A
–600
A
600
A
2.2 A
3
2.195759 A
2.204241 A
2.196199 A
2.203801 A
Table 26.
AC Current Verification Limits
Calibrator Output
Range (A)
1-Year Limits
24-Hour Limits
Amplitude
Frequency
Lower
Upper
Lower
Upper
200
A
1 kHz
10 m
193.4
A
206.6
A
194.4
A
205.6
A
1 mA
1 kHz
10 m
0.991 mA
1.009 mA
0.992 mA
1.008 mA
10 mA
1 kHz
10 m
9.964 mA
10.036 mA
9.965 mA
10.035 mA
2 mA
1 kHz
100 m
1.934 mA
2.066 mA
1.944 mA
2.056 mA
10 mA
1 kHz
100 m
9.91 mA
10.09 mA
9.92 mA
10.08 mA
100 mA
1 kHz
100 m
99.64 mA
100.36 mA
99.65 mA
100.35 mA
10 mA
1 kHz
500 m
9.67 mA
10.33 mA
9.72 mA
10.28 mA
50 mA
1 kHz
500 m
49.55 mA
50.45 mA
49.6 mA
50.4 mA
500 mA
1 kHz
500 m
498.2 mA
501.8 mA
498.25 mA
501.75 mA
60 mA
1 kHz
3
58.02 mA
61.98 mA
58.32 mA
61.68 mA
300 mA
1 kHz
3
297.3 mA
302.7 mA
297.6 mA
302.4 mA
2.2 A
1 kHz
3
2.1916 A
2.2084 A
2.1919 A
2.2081 A
Table 25.
DC Current Verification Limits (Continued)
Calibrator Amplitude
Range (A)
1-Year Limits
24-Hour Limits
Lower
Upper
Lower
Upper