NI 4065 Calibration Procedure
80
ni.com
17. Call the niDMM Cal Adjust Offset VI.
18. Output 2.2 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
19. Call the niDMM Cal Adjust Gain VI.
20. Output –2.2 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
21. Call the niDMM Cal Adjust Gain VI.
22. Output 1 A on the calibrator.
Allow the calibrator output to settle
before proceeding
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with
the following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
0.1
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
3
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
2.2
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with the
following parameters:
Instrument_Handle
:
Cal Session
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
3
I
nput_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
–
2.2