NI 4065 Calibration Procedure
50
ni.com
5.
Call the niDMM Config Measurement VI. The niDMM Config
Measurement VI is polymorphic. To change the selected instance of
this polymorphic VI, right-click the VI and choose
Select Type
followed by
Resolution in Digits
.
6.
Call the niDMM Read VI to configure the NI 4065 for a current mode
before applying current.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Resolution_Digits
:
6.5
Measurement_Function
:
NIDMM_VAL_AC_CURRENT
Range
:
0.01
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
: The instrument
handle from
niDMM_init
Maximum_Time
: –
1