Equipment Environment and Safety Specifications
Appendix
19 - 2
Passport V Operating Instructions
Transport
Complies with the requirements of ISTA transport test procedure 2A (for shipping cargos in
containers).
Noise Level
• The patient monitor complies with the requirements of ISO 3744 in normal operation.
• Temperature controlled fan is used.
Below and equal to 27 degree ambient temperature, noise
≤
35 dBA measured at
a distance of one meter from the device.
• Below and equal to 33 degree ambient temperature, noise
≤
40 dBA measured at a
distance of one meter from the device.
• Below and equal to 40 degree ambient temperature, noise
≤
46 dBA measured at a
distance of one meter from the device.
• Conditions: NIBP pump and valves not operational, local printer not operational, CO
2
not operational.
• SPL range does not include patient heartbeat tone, key-pressing tone and alarm tone.