IBP Description
IBP Monitoring (optional)
9 - 2
Passport V Operating Instructions
The Default labels: Art, CVP.
The pressure labels for each channel cannot be the same. That is, if a pressure label is
selected in one channel, it will not appear among the available label options for the other
channel.
P1 and P2 can be set to Pulsatile or Mean. The default setup is Pulsatile.
• When 'Pulsatile' is selected, the channel displays data in pulsatile pressure format
SYS/DIA (MEAN) and the waveform scale defaults to 0-160 mmHg. The measured
systolic value is displayed.
• When 'Mean' is selected, the channel displays data in non-pulsatile pressure format
MEAN (SYS/DIA) and the waveform scale defaults to 0-20 mmHg. The measured mean
pressure value is displayed.
• For all other pressure labels, the data will be displayed in a fixed format, depending on
the Measure Pressure selection in the IBP Menu. If “Pulsatile” is selected, the measured
systolic (SYS/DIA (MEAN)) is displayed. If “Mean” is selected, the measured mean
pressure value (MEAN (SYS/DIA)) is displayed.
CVP
Central Venous Pressure
ICP
Intracranial Pressure
LA
Left Atrial Pressure
RA
Right Atrial Pressure
P1, P2
Not specified pressure
labels
LABEL
DEFINITION