OxyCRG
Trends
15 - 8
Passport V Operating Instructions
• The x-axis is time, and the y-axis is parameters.
• The most recent trend data is shown at the rightmost end.
• The time marker is exact to minute.
• Parameters in OxyCRG screen are displayed in the order of HR, SpO2, Resp Wave,
Temp and IBP1.
• The alarmed parameters are marked with different color under the OxyCRG graph to
match the alarm level.
• 1 level alarms: Red
• 2 level alarms: Yellow
• The parameters' trend graph is the same color as the parameter's color.
The rules for displaying parameter values are as follows:
• The color of the alarmed parameter is the same as the color of the alarm priority.
• one level alarms: Red
• two level alarms: Yellow
• The invalid parameter data is shown as “– – –“ in the parameter area of the OxyCRG
screen.
• The IBP parameter value in which the IBP channel isn't zeroed shall be displayed as
“– – –“ in the parameter area of the OxyCRG screen.
The rules for displaying events are as follows:
Manually marked events are marked in green in the Event row.
• Alarm events are marked in different colors in the Event row to match the alarm priority.
• 1 level alarm: red
• 2 level alarm: yellow
• If multiple events happen during a time interval, the color of the event with the highest
priority is shown.
• The order of event priority is: manually marked event, one level alarm, two level alarm.
Viewing Trend Data
The start time for OxyCRG trends is related to the way the user accesses the OxyCRG trends:
• If the user enters the OxyCRG trends by pressing the Trend key 4 times in normal screen,
the start time for the OxyCRG trend display is accurate to the minute of the time when the
user accesses the OxyCRG trends. (The method for obtaining the integrated start time of
the OxyCRG trends is the same with that of the List trends.)
• If the user accesses the OxyCRG trend by pressing the OxyCRG button in the Quick
Trend screen, the cursor time of OxyCRG is accurate to minute of the time of most recent
trend data in the Quick Trends.
• If the user accesses OxyCRG trends by pressing the OxyCRG button in the List Trend
screen, the cursor time of OxyCRG trends is accurate to minute of the time of rightmost
column in the List Trend.
• If the user accesses OxyCRG trends by pressing the OxyCRG button in the Graph Trend
screen, the cursor time of OxyCRG trends is accurate to minute of the time to which the
cursor point in the Graph Trends.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...