Quick Trends
Trends
15 - 2
Passport V Operating Instructions
15.2
Quick Trends
The
Quick Trend
display allows the user to view an abbreviated numeric listing of HR,
SpO
2
and NIBP data only. To access this display from the normal monitoring screen, press
the
TRENDS
key once. This display may also be accessed from the other trend displays via
a menu choice. A maximum of 9 time-stamped entries may be stored. When the maximum
number of entries has been reached, the oldest entry will be deleted from the trend record in
order to allow storage of a new entry.
FIGURE 15-1
Quick Trends
Clearing Trend Data
To manually clear all trend data, including Graphic Trends, choose
Clear Trends
from the
menu. A confirmation prompt will appear. Once cleared, the data cannot be restored.
• All trend data is automatically cleared when the patient is discharged from the monitor.
• All trend data is also cleared if the monitor’s displayed time or date is changed.
Removing the Quick Trend Display
The
Quick Trend
display does not automatically time-out and must be manually removed to
return to the normal waveform display. To remove the
Quick Trend
display, choose
Normal Screen
from the menu, or press the
NORMAL SCREEN
key.