Passport V Operating Instructions
3 - 9
Patient Management
Remote View
4. Remote Silence / Remote Silence All
— this selection silences the audio portion of
a remote alarm for the duration that is selected from the
Remote Silence For /
Remote Silence All For
list in the
Remote View Setup Menu
. The factory default
is 2 minutes. While the audio alarm is silenced:
• The visual alarm indicators remain displayed.
• The message
ALARMS SILENCED FOR X:XX mins
is displayed in the message
area of the
Remote View Menu
. The
X:XX
in the message is a digital timer for the
silence time remaining.
• If a new remote alarm occurs during this time, the current alarm will remain silenced
while its digital timer continues to count down, and the new alarm tone will sound.
• An Alarm Silence icon is displayed in the top left of the Remote View window.
• If
Remote Silence
is selected again, the digital timer is reset.
NOTE:
The Remote Silence selection silences only the alarms
indicated in the Remote View Menu. It does not silence the
primary patient's alarms or alarm sounds at the remote
location.
5. Remote Print
– Select to request a remote print.
6. Remote Silence All
— this selection silences the audio alarm tones of all remote
alarms for the duration that is selected from the
Remote Silence All For
list in the
Remote View Setup Menu
. The factory default is 2 minutes. While the audio alarms
are silenced:
• the visual alarm indicators remain displayed
• the message
All Alarms Silenced For X:XX mins
is displayed in the message
area of the
Remote View Menu
. The
X:XX
in the message is a digital timer for the
silence time remaining
NOTE:
If “Permanent” is selected from the Remote Silence All For
list, “All Alarms Silenced Permanently” is displayed in the
message area of the Remote View Menu.
• an Alarm Silence icon (a crossed bell) is displayed in the numeric tiles and in the
“Numeric Data Area”
If
Remote Silence All
is selected again, the audio alarm tones are re-enabled.
NOTE:
The Remote Silence All selection silences only the alarms
indicated in the Remote View Menu. It does not silence the
primary patient's alarms or alarm sounds at the remote
location.
7. Remote Print
— this selection sends a print request to the remote monitor. The printout
will be as configured in the
Print Setup
menu of the remote monitor.
NOTE:
When a Remote Print is requested, there is no indication that
the printout was completed as requested. The user should
verify the successful print at the printer.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...