Temperature Troubleshooting
Temperature Monitoring
8 - 8
Passport V Operating Instructions
8.4
Temperature Troubleshooting
\
MESSAGE/PROBLEM
REASON
SOLUTION
Temperature Probes not
Working
Poor contact from probes to body
Check the body surface contact
at the probe tip.
Reposition or apply
thermoconductive gel.
Temperature not displayed
Improper display setup
Check
Display Setup
in
Monitor Setup Menu
and
change as desired.
Cable not plugged in
Check the cable.