Passport V Operating Instructions
7 - 9
NIBP Monitoring
NIBP Measurements
1.
Select the
Interval Mode
or the
Timer Mode
in the
Monitor Setup Menu
〉
Advanced Setup
.
2.
Select the
NIBP
tile (or
Parameters
)
〉
NIBP
〉
Interval
. Rotate the Navigator Knob
to the desire selection. The selections are:
OFF
,
STAT
,
1
,
2
,
3
,
5
,
10
,
15
,
20
,
30
,
1
hr
,
2 hr
and
4 hours
.
3.
Press the
Start
key to begin taking interval measurements.
NOTE:
If the monitor is in the interval mode when it is turned ON,
no measurement will be taken until the Start key is pressed.
Automatic Adjustment in the Interval Mode
In the
Interval Mode
, the unit adjusts the inflation pressure according to the previous
reading of the systolic pressure. After the first measurement in the timer mode, the inflation
pressure is the previous sy50 mmHg in the
Adult Mode
or
Pediatric Mode
and
+40 mmHg in the
Neonate Mode
.
7.4.3
Suspension of NIBP Measurements
1.
Press
STOP
to suspend an automatically timed measurement sequence or to end a
measurement cycle already in progress (deflate cuff).
2.
Press
START
to take an immediate measurement and resume a suspended timed
measurement sequence.
NOTE:
You can press STOP at any time to postpone a scheduled
measurement or to terminate a measurement cycle already
in progress.
CAUTION:
7.4.4
Retry measurement
If an NIBP measurement fails, the measurement is retried up to three times. The message
"NIBP: Retry" is displayed after the cuff deflates. If the measurement fails all three retries, the
Display shows the Sys, Dia, and Mean values as "XXX".
7.4.5
Reset
Reset can occur as NIBP module error(e.g. communication error). If the cuff is being inflated,
it will deflate as soon as it receives the reset signal.
When the module resets, the prompt message "NIBP: Resetting..." is displayed in the
message area B.
7.4.6
NIBP Pressure Limit Fail Safe
If the cuff is over-pressurized, the cuff will automatically vent to atmosphere.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...