Warnings
Introduction
WARNING: Use of the Passport V in the vicinity of explosive anesthetics
and in the presence of electromagnetic interference or power
overloads caused by electrosurgical or diathermy instruments
could create a potential hazard or could damage the monitor.
WARNING: The use of gas sampling accessories in Gas Module 3 other than
those specified may cause significant measurement errors and
patient risk.
WARNING: Use of accessories, transducers and cables other than those
specified in the manual may result in increased Electromagnetic
Emissions or decreased Electromagnetic Immunity of the Gas
Module 3.
WARNING: With the exception of stacking on a Gas Module with the
appropriate mounting brackets, the Passport V should not be
used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the Passport V and Gas Module
should be observed to verify normal operation in the
configuration in which they will be used.
WARNING: If the water trap breaks or becomes damaged during
operation, there is a risk that bacteria and/or mucus may
contaminate the Gas Module.
WARNING: The Gas Module must not be used with flammable anesthetic
agents.
WARNING: SpO
2
sensors, SpO
2
accessories, and temperature probes
should be disposed of in accordance with local regulations.
WARNING: Shut down the monitor and disconnect all power cords from the
outlet before cleaning.
WARNING: Connect only DRYLINE
™
gas sampling lines to the water trap.
Note that there may be other compatible tubes present that
must not be used (e.g. IV lines).
WARNING: Do not use DRYLINE
™
Neonatal sampling lines (blue Luer lock
nuts) with DRYLINE
™
Adult/Pediatric water traps. Do not use
DRYLINE
™
Adult/Pediatric sampling lines (colorless Luer lock
nuts) with DRYLINE
™
Neonatal water traps. These
configurations could result in incorrect measurement data.
WARNING: Do not use Adult/Pediatric type water traps and/or sampling
lines with neonates to avoid high sampling flow.
WARNING: The contents of the Gas Module water trap should be handled
as a potential infection hazard. The water trap, sampling line
and airway adapter should be disposed of in accordance with
local regulations for contaminated and biologically hazardous
items.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...