Performance Characteristics
Appendix
19 - 18
Passport V Operating Instructions
Masimo SET
®
Reference notes
The Masimo SET MS-3 pulse oximeter with LNOP
®
•Adt sensors has been validated for no
motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia
studies in the range of 70% to 100% SpO
2
against a laboratory co-oximeter and ECG
monitor. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population.
The Masimo SET MS-3 pulse oximeter with LNOP
®
•Adt sensors has been validated for
motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia
studies while performing rubbing and tapping motions at 2 to 4 Hz. At an amplitude of 1 to
2 cm and non-repetitive motion between 1 to 5 Hz. At an amplitude of 2 to 3 cm in induced
hypoxia studies in the range of 70% to 100% SpO
2
against a laboratory co-oximeter and
ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population.
The Masimo SET MS-3 pulse oximeter with LNOP
®
•Neo and Neo•Pt sensors has been
validated for neonatal motion accuracy in human blood studies on neonates while moving
the neonate's foot at 2 to 4 Hz at an amplitude of 1 to 2 cm against a laboratory co-oximeter
and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus
one standard deviation encompasses 68% of the population.
The Masimo SET MS-3 pulse oximeter has been validated for low perfusion accuracy in
bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal
strengths of greater then 0.02% and a % transmission of greater than 5% for saturation's
ranging from 70 to 100%. This variation equals plus or minus one standard deviation. Plus
or minus on standard deviation encompasses 68% of the population.
Low Perfusion Performance
LOW PERFUSION CONDITIONS
PULSE
AMPLITUDE
%
TRANSMISSION
SATURATION
ACCURACY
PULSE RATE
ACCURACY
> 0.02%
> 5%
± 2 digits
± 3 digits
Pulse Rate Range and Accuracy
ACCURACY
PATIENT SIZE
PULSE RATE
RANGE
NO MOTION
CONDITIONS
DURING
MOTION
CONDITIONS
Adult / Pediatric / Neonate
25 to 240 bpm
± 3 digits
± 5 digits
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...