Passport V Operating Instructions
19 - 27
Appendix
Performance Characteristics
Rise times (t10–90%)
3
@120 ml/min
• CO
2
250 ms (fall time 200 ms)
• N
2
O 250
ms
• O
2
600
ms
• HAL, ISO, SEV, DES 300 ms
• ENF
350 ms
Delay time
4
(t0—10%)
< 4 s
Identification Dual
agent
Primary agent ID threshold
5
0.15% (0.4% during ISO accuracy mode)
Secondary agent ID threshold
5
0.3% (0.5% during ISO accuracy mode) or 5% REL
(10% REL for Isoflurane) of primary agent if primary
agent >10%
Agent ID time
Three breaths - Typically less than 10 s
Display Update Rate Breath-by-Breath
Compensation
The Gas Module 3 is equipped with automatic barometric
pressure compensation.
Main Fuse
2x T0.8A 250V
1
Includes interference from other gases.
2
The step rise time specification at 200 ml/min sample flow includes DRYLINE• Water Trap, Adult/Pediatric
and DRYLINE• Sampling Line, Adult/Pediatric 2.5 m.
3
The step rise time specification at 120 ml/min sample flow includes DRYLINE• Water Trap, Neonate and
DRYLINE• Sampling Line, Neonate 2.5 m.
4
The delay time specification is valid both for 120 ml/min sample flow (using DRYLINE• Water Trap, Neo-
nate and DRYLINE• Sampling Line, Neonate 2.5 m) and for 200 ml/min sample flow (using DRYLINE•
Water Trap, Adult/Pediatric and DRYLINE• Sampling Line, Adult/Pediatric 2.5 m).
5
For HAL, add 0.1% ABS to threshold values.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...