Passport V Operating Instructions
14 - 5
Messages
Technical Alarm Messages
14.2.3
TEMP Alert Message
14.2.4
SpO
2
Alert Message
14.2.5
NIBP Alert Message
Resp Disturbed
3
Patient Movement
External Interference.
Check patient
For impedance - Check
Electrode Contacts /
reposition electrodes/cable
Check environment for source
of interference.
Resp Rate Overrange
3
Respiration Outside
of measureable range
Check patient.
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
Temp Overrange
3
Temperature Outside
of measureable range
Check patient,
Check sensor
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
SpO
2
: Sensor Off
3
Sensor Off
Check Patient, Apply Sensor
Note that the alarm is only
active after the sensor has
been on a patient.
SpO
2
: Overrange
3
SpO
2
outside of
measurable range
Check patient
SpO
2
: PR Overrange
3
PR outside of
measurable range
Check patient
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
NIBP: Systolic Overrange
3
NIBP systolic pressure
exceeds the
measurement range.
Check Patient.
Ensure that the measured
pressure label is within the
measurement range.
NIBP: Diastolic Overrange
3
NIBP diastolic
pressure exceeds the
measurement range.
Check Patient.
Ensure that the measured
pressure label is within the
measurement range.
NIBP: Mean Overrange
3
NIBP mean pressure
exceeds the
measurement range.
Check Patient.
Ensure that the measured
pressure label is within the
measurement range.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...