Passport V Operating Instructions
16 - 3
Printing
Print Setup Menu
Print on Alarm
If a waveform is initiated by an alarm violation, then the local printer will initiate a 16-second
or 32-second waveform strip. The print length (16 or 32-seconds) can be set in the Print
Setup menu. To activate this feature, from the front panel, select
Alarms: Limits
〉
Print On
Alarm
〉
Active
. Then select each parameter for which the feature should be enabled.
During an alarm printout, the data for the parameter which violated the alarm will be marked
with an asterisk. The alarm trigger for the printout is indicated on the strip with the name of
the violated alarm.
Interval Printing
To print waveform strips at regular intervals, go to the
Print Setup Menu,
highlight
Interval
and select a time interval to print. The selections are
Off, 1 min, 5 min, 10
min, 15 min, 20 min, 30 min, 1 hr
and
2 hr
. The source of the printout is indicated on
the strip as
AUTO REALTIME REPORT
.
Trend Printing
To print
Quick Trend
,
List Trend
,
OxyCRG,
or
Graphic Trend
data to the local printer,
press the
Print Trend
key while the trend is displayed. To abort printing, press the
Cont
ECG
,
Strip
, or
Print Trend
key.
Quick Trend and List Trend print digital data information while Graphic Trend print graphic
representations of the data in the respective menus. Quick Trend entries are only recorded
upon NIBP readings.
FIGURE 16-2
Graphic Trend Format
FIGURE 16-3
List Trend Format
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...