Technical Alarm Messages
Messages
14 - 6
Passport V Operating Instructions
14.2.6
IBP Alert Message
14.2.7
CO
2
Alert Message
14.2.8
GAS Alert Message
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
<IBP Label>: Systolic
Overrange
3
IBP systolic pressure
exceeds the
measurement range.
Ensure that the measured
pressure label is within the
measurement range.
<IBP Label>: Diastolic
Overrange
3
IBP diastolic pressure
exceeds the
measurement range.
Ensure that the measured
pressure label is within the
measurement range.
<IBP Label>: Mean
Overrange
3
IBP mean pressure
exceeds the
measurement range.
Ensure that the measured
pressure label is within the
measurement range.
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
CO
2
: ETCO
2
Overrange
3
Et CO
2
outside of
measurable range
Check patient
CO
2
: InspCO
2
Overrange
3
Insp CO
2
outside of
measureable range
Check patient
CO
2
: Resp Rate
Overrange
3
The measured RR
value exceeds the
measurement range.
Check Patient.
Check Airway.
ALARM MESSAGE
PRIORITY
REASON
SOLUTION
GM: ETCO
2
Overrange
3
The measured
ExpCO
2
value
exceeds the
measurement range.
Check Patient.
Check Airway.
GM: InspCO
2
Overrange
3
The measured InsCO
2
value exceeds the
measurement range.
Check Patient.
Check Airway.
GM: Resp Rate
Overrange
3
The measured RR
value exceeds the
measurement range.
Check Patient.
Check Airway.
GM: Disconnected
2
Appears when the
Passport V
cannot
detect signals being
sent by the Gas
Module.
Ensure Gas Module is turned
on and interface cable is
properly connected. If
problem persists, contact
Technical Support.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...