Passport V Operating Instructions
13 - 9
Alarms
Alarms
Alarms also can be verified by creating an alarm condition on the
Passport V
and
verifying the corresponding alarm indicators are present on the monitor.
Alarm Delay
• Apnea Alarm Delay settings are: Adult/Pedatric: 10/15/20/25/30/35/40/45/50/
55 seconds, 1 min
Neonate: 10/15/20 sec
• ST Alarm Delay settings are:
30 seconds/45 seconds/ 1 min/90 seconds/2 min/3 min.
• Alarm delay settings are:
None/1s/2s/3s/4s/5s/6s/7s/8s.
Audio alarm can be delayed for a predetermined time period, with the exception of Apnea,
ST alarm, lethal Arrhythmia alarms, SpO
2
Desat and Insp O
2
shortage.
When alarm is triggered, the alarm tone will not be presented, but the visual signal will be
presented during delay time.
13.1.6
Alarm Troubleshooting
MESSAGE/PROBLEM
REASON
SOLUTION
High or Low or No Alarm
Sound
Alarm limits not set
Go to
Alarm Setup
and adjust
alarms
Alarm
Silence All, On
time has
not expired
Press
SILENCE ALL
to reactivate
alarms
No Arrhythmia
Alarm Sound
Arrhythmia option not installed
Call Sales Rep to purchase option
Arrhythmia Alarms off
Go to
Monitor Setup
〉
Advanced Setup
to activate
alarm
Monitor is in learning mode
Wait until learning is concluded
and monitor patient closely
Alarms continue to Sound
despite pressing
SILENCE
More than one alarm is active
Press
SILENCE
or
SILENCE ALL
key to silence
Check Patient
No Alarm printout with
Alarm violation
Print on Alarm
is set to
Suspend
Go to
Print Menu
and set
on Alarm
to
Active
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...