Passport V Operating Instructions
19 - 25
Appendix
Performance Characteristics
19.8.9
Temperature Performance Characteristics
The
Passport V
is capable of providing temperature measurements from one channel for
400 series probe.
Measuring range: 0 - 50°C (32 -122°F)
Resolution: 0.1°C (0.1°F)
Scale: °C, °F
Tabular Trend: Yes
Graphic Trend: Yes
Clinical Common Range Default of Graphic Trend: 32 - 43°C (90 - 110°F)
Quantitative Influence of
Interference Gas
Interference gas
Concentration
Level (%)
Quantitative
Influence
When measuring
under 0~40mmHg,
additional error
introduced due to
gas interference
N
2
O
60
±1mmHg
HAL
4
±1mmHg
SEV
5
±1mmHg
ISO
5
±1mmHg
ENF
5
±1mmHg
DES
15
±2mmHg
Helium
Unspecified
Xenon
Unspecified
Start Time:
30 s (typical) 60s (maximum)
Warm-up Time:
1 min
Flowrate:
Options are 70ml/min, 100ml/min, 120ml/min and
150ml/min.
Flowrate Accuracy:
±15% or ±15ml/min, whichever is greater