NIBP Measurements
NIBP Monitoring
7 - 8
Passport V Operating Instructions
5.
If necessary, change the initial cuff inflation pressure through the
NIBP Menu
. Initial
cuff inflation pressures depend on the
patient size
setting. The values for cuff inflation
are identified in the following table:
6.
Press
START
to begin an NIBP measurement.
CAUTION:
Blood pressure cuffs must be used with the manufacturer’s
correct and approved hoses.
CAUTION:
Please consult a physician for interpretation of blood
pressure measurements.
CAUTION:
NOTE:
Inflate the cuff only after proper application to the patient’s
limb. Cuff damage can result if the cuff is left unwrapped
and then inflated.
NOTE:
Changing the patient size, during measurement will stop the
current measurement and the cuff shall be deflated.
7.
The cuff begins to inflate to the selected cuff pressure. After reaching the selected value
the cuff begins to slowly deflate and the
Passport V
collects oscillometric pulsations.
8.
If the initial cuff inflation is found to be inadequate, the unit retries with a higher inflation
pressure (+50 mmHg in adult or pediatric mode; +40 mmHg in neonate mode).
9.
The patient should remain still to avoid the introduction of unnecessary motion artifact.
After the cuff pressure drops below the diastolic pressure, the results of the measurement
are displayed.
10.
If
NIBP End Tone
is set to
On
in the
NIBP Menu
, a double tone will sound to
indicate a successful NIBP measurement.
The NIBP measurement will be automatically removed from the display after a predetermined
time interval. The NIBP timeout interval is 15 minutes by default and can be set to a different
value through the
Installation Menu
.
7.4.2
Automatic Interval NIBP Measurements
There are two modes available for automatic NIBP measurements. They are the
Interval
Mode
and the
Timer Mode
. The
Interval Mode
allows you to set the interval between
measurements. For example, if the interval is set to 10 minutes and the
START
key is pressed
at 10:12, the measurements will be taken at 10:12, 10:22, 10:32, etc. The
Timer Mode
allows you to set an interval that is synchronized with the real time clock. For example, if the
timer is set to 30 and the
START
key is pressed at 10:12, the measurements will be taken at
10:12, 10:30, 11:00, 11:30, etc.
PATIENT SIZE
SETTING
INITIAL CUFF
INFLATION VALUES
DEFAULT
SETTING
MAXIMUM
INFLATION VALUES
Adult
100 - 280 mmHg
180 mmHg
295 mmHg
Pediatric
60 - 180 mmHg
140 mmHg
190 mmHg
Neonate
40 - 120 mmHg
100 mmHg
145 mmHg
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...