background image

 

1

Eickemeyer 

 
 
 

MAGIC 500   

Ultrasonic Diagnostic Instruments 

 

User’s Manual   

 

 

 
 
 
 
 
 
 
 
 

 
 

 

 

 

 

 

 

 

 

 

Information for Users 

Users shall carefully read through this manual and fully understand the text before operating the equipment. This manual shall 

be placed easy of access for future reference; otherwise it may cause instrument damage or person injury. 

 

Summary of Contents for MAGIC 500

Page 1: ...r s Manual Information for Users Users shall carefully read through this manual and fully understand the text before operating the equipment This manual shall be placed easy of access for future reference otherwise it may cause instrument damage or person injury ...

Page 2: ...responsibilities of implied marketability or merchantability applicable to the product for a particular use Eickemeyer shall be responsible for the safety reliability and proper performance of this equipment on condition that the assembly extension readjustment modification or service of the equipment herewith related is done by technical personnel authorized or approved by Eickemeyer Medizintechn...

Page 3: ...uipment To obtain the right to return the goods contact local dealer of Eickemeyer Company indicating product serial number on the packing case or parts serial number on the part product model and reason to return the product through the dealer to return the products or parts Manufacturer s Information EICKEMEYER MedizintechnikfürTierärzteKG Eltastraße8 78532Tuttlingen Germany email info eickemeye...

Page 4: ... 96 EC The device accessories and the packaging have to be disposed of waste correctly at the end of the usage Please follow Local Ordinances or Regulations for disposal F 3 Contents C h a p t e r O n e Te c h n i c a l S p e c i f i c a t i o n s 1 C h a p t e r T w o S y s t e m O u t l i n e a n d S t r u c t u r e 2 2 1 S t r u c t u r e c o m p o s i t i o n o f t h e i n s t r u m e n t 2 2 ...

Page 5: ... s t m e n t 1 0 6 5 4 D y n a m i c R a n g e A d j u s t m e n t 1 0 6 6 Selection of Magnification Frequency Frame Correlation Image Post process Edge Enhancement Focus Number L o c a l Z o o m a n d D e p t h H e i g h t e n 1 0 6 6 1 M a g n i f i c a t i o n S e l e c t i o n 1 0 6 6 2 F r e q u e n c y S e l e c t i o n 1 0 6 6 3 F r a m e C o r r e l a t i o n S e l e c t i o n 1 1 6 6 4 I...

Page 6: ...4 S y s t e m m a i n t e n a n c e 2 2 C h a p t e r Te n S t o r a g e a n d T r a n s p o r t a t i o n 2 3 C h a p t e r E l e v e n S a f e t y C l a s s i f i c a t i o n 2 3 A p p e n d i x S y s t e m B l o c k D i a g r a m 2 4 Chapter One Technical Specifications 1 1 Technical Data 1 Grey scale 256 2 Monitor 5 7 LCD 3 Main power supply 100 240V frequency 50 60Hz 4 Rating power 40VA 5 Mai...

Page 7: ...tch map 2 3 Parts of the probe Take 6 5MHz animal transrectal linear array probe for example Fig Parts name of 6 5MHz animal transrectal linear array probe Name Function 1 Acoustic lens Toconvertelectricsignaltoultrasonicsignalbasedonprincipleofconversepiezoelectric effect Theultrasonicsignal afterenteringthehumanbody isreflectedasechowaveand convertedtoelectricsignalagain Theacousticlensisonthepr...

Page 8: ...high frequency linear array probe 3 5 MHz convex array probe 2 Video printer P93W S 3 Mouse Chapter Four Operation Condition 4 1 Power supply Mains voltage 100 240 10 Mains frequency 50 60Hz 1Hz 4 2 Operation Environment Ambient temperature 10 40 Relative humidity 30 75 without condensation Atmospheric pressure 700hPa 1060hPa 4 3 Storage and Transport Ambient temperature 20 55 Relative humidity 30...

Page 9: ...faults Warning 1 When instrument works abnormally do stop working turn off the power and check the reason then contacts the EICKEMEYER Company about it 2 Turn off power and pull out of the plug from socket after each ultrasonic diagnostic operation 3 It is forbidden to drag and press the power and probe cables emphatically regularly inspect whether there is pull apart and bareness if there is the ...

Page 10: ...ng or using the probe make sure that the probe connecting cable and connector are in normal condition free of cracks or drop Use of defective probe may cause electric shock Warning 1 The Eickemeyer ultrasonic probe shall be connected to the dedicated Eickemeyer ultrasonic system only Select proper probe model according to the relevant instructions of ultrasonic diagnostic system 2 Check the ultras...

Page 11: ...core power supply It connects with the protective earth line when power plug inserts into its socket 5 6 Video printer installation 1 Connect the equipotential terminal of the video printer to the earthing 2 Connect one end of the video connection line to the video printer and the other end to the video output interface on the left of main unit 3 Insert one end of power plug jack of the video prin...

Page 12: ... machine 1 No abnormal voice strange smell and overheating appear 2 Check the machine to ensure a normal start up The power indication light is on and startup picture is shown on the screen The machine will be then automatically set in B mode 3 Check the acoustic lens for abnormal heat when the probe is in use This can be done by hand touching the probe to feel the temperature of the lens 4 Check ...

Page 13: ... 192 224 and 255 5 Press key to quit setting interface 6 2 5 Hospital Name Setting 1 Press keytoentermain menuandpressdirectionkeys tomovecursorto Preset 2 Press direction keys to enter setting interface 3 Press direction keys to move symbol to point to Hospital 4 Press key the cursor is located above NAME at the same time characters input menu will be shown at the bottom of the screen Caps 0 1 2 ...

Page 14: ...nd a single framed B mode image is displayed 6 4 2 B B Mode 1 In real time mode press key to enter B B mode 2 B B image switch Press key to enter main menu and press direction keys to move the cursor to B B Mode and then press direction keys to switch image display Or right click mouse to switch image display the selected image is activated and the other one is frozen 3 In real time mode press key...

Page 15: ...tion Image Post process Edge Enhancement Focus Number Local Zoom and Depth Heighten 6 6 1 Magnification Selection In real time mode press key to enter main menu and move cursor to Zoom and then press direction keys to choose eight kinds of magnification 6 6 2 Frequency Selection In real time mode press key to enter main menu and press direction keys to move cursor to Freq in the display area Press...

Page 16: ...on keys to realize image vertical reverse 6 7 3 Image Negative In real time mode press key to enter main menu and press direction keys to move cursor to Image POS and then press direction keys to realize negative function 6 7 4 Color Selection In real time mode press key to enter main menu and press direction keys to move cursor to COLOR in the display area and then press direction keys to realize...

Page 17: ...lect the image code need to be saved in the U disk such as choose 004 5 Press key to appear Wait prompt After the prompt disappear the current image is saved in the frame for coded 004 The current image is saved under the specified folder in the U disk the folder name is patient number ID which is made up of underline letters or numbers without spaces file name is the selected saving code If user ...

Page 18: ...s cine loop key until enter PAUSE status 2 Press key LOCAL U Disk Mass Dump prompt display at the bottom of the screen 3 Press key Dumping prompt appears images are mass dumped under the DUMP folder in the U disk and the number behind PAUSE character is the number of folder name of mass dumping After finishing mass dumping return to PAUSE status 4 Press key to quit mass dumping status and press ke...

Page 19: ...nes 6 After inputting ID press key and press direction keys to move cursor to image area and press key again to input according to Step 4 7 If need modify the content continuously press key twice to quit annotation status press direction keys to select Clear at last press key to clear all noted marks and retype 8 Press key to quit 6 14 Case report processing center V3 1 software optional 6 14 1 Fu...

Page 20: ...ace press key again the cursor will show or click left mouse the cursor will show 3 Press direction keys or operate mouse to move the mark to desired position press key or click left mouse to set the mark position as the start of the measurement or click right mouse to set the mark position as the start of the measurement 4 Press direction keys or operate mouse to move the mark to the end of the m...

Page 21: ...surement and Calculation items Obstetric tables of the system including 1 BOVINE 2 EQUINE 3 OVINE 4 CANINE 5 FELINE 6 GOAT 7 LLAMA 8 SWINE which is reference for doctor 8 2 Measurement of Gestational Age GA and Estimated Date of Confinement EDC Follow the steps below 1 In B B B mode freeze the desired image press direction keys to choose OB in the display area 2 Press key to display the obstetric ...

Page 22: ...ear all marks and data and quit the measurement status 8 3 Obstetric report In frozen status press key to enter main menu and press direction keys to move the cursor to the position Report Press direction keys to display obstetric report Press key to exit obstetric report status 8 4 Measurement items 1 Items measurable in B mode distance circumference area volume gestational age GA and estimated d...

Page 23: ...n off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any water roof device in the system Do not splash any water or liquor into the system otherwise it will cause malfunction or electric shock Attention 1 Probe without cleaning disinfection or sterilization may be...

Page 24: ...h dipped in moderate detergent and then air dry it 9 1 2 Use and maintenance for the charging battery 1 Only use charger and corresponding model of battery provided by Eickemeyer Company 2 The output port of adapter plugs into the input port of charger to charge the minimum charging time is 3 hours and the full charged battery can be used for 1 5 2 hours 3 The fore three times battery charging sho...

Page 25: ...her with one off battery such as dry battery or different capability or different model or different brand battery which will lead to the leakage explosion or fire 13 Do not use the abnormal battery with particular smell or abnormal heat or distortion or turn colors or abnormal phenomena which will lead to the leakage explosion or fire 14 Do stop the charge and pull out the battery from the charge...

Page 26: ... the machine 1 Check power supply 2 Check power cable and connections 3 Check power adapter Character and grey scale are displayed but no ultrasonic image on the screen Probe is not properly connected Turn off the power and reconnect the probe Intermittent stripe snow or far field interference appears on screen 1 Check power supply spark interference present 2 Check environment source of interfere...

Page 27: ...ell ventilated and free of direct sunlight and corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Eleven Safety Classification 1 Classified according to electric shoc...

Page 28: ...28 Appendix System Block Diagram Probe Transmission Reception Keyboard Microcomputer control system Clock Line field sync signal Collating DSC Process Signal synthesis Display ...

Page 29: ...nik für Tierärzte KG Eltastraße 8 78532 Tuttlingen Germany tel 49 0 7461 96580 0 fax 49 0 7461 96580 91 mail export eickemeyer de web www eickemeyer de Information contained in this manual is subject to change without further notice ...