Please carefully read through and fully understand the safety cautions before moving and placing the system.
1. Unpack the instrument case and check the goods for its completeness according to the packing list furnished.
2. Place the instrument on a stable and leveled position.
3. Leave adequate space of 20 centimeters as minimum from rear, left and right side of the instrument.
5.2 Probe installation
5.2.1 Probe connection
Insert the probe connector into the probe socket at the back of the main unit;
5.2.2 Probe disconnection
Turn off the system, and use probe screwdriver to remove the probe with figure’s method and instructions.
Socket screwdriver
According to arrow direction,
gently press socket screwdriver
and remove the probe.
5.3 Shutter release installation
Install shutter release as shown in the figure.
Attention:
Adequate space from rear, left and right side of the machine shall be reserved,
or the machine may malfunction under excessive heat inside the enclosure.
Warning:
Before connecting or using the probe, make sure that the probe, connecting
cable and connector are in normal condition (free of cracks or drop). Use of defective
probe may cause electric shock.
Warning:
1. The Eickemeyer ultrasonic probe shall be connected to the dedicated Eickemeyer ultrasonic
system only. Select proper probe model according to the relevant instructions of ultrasonic
diagnostic system.
2. Check the ultrasonic probe and connecting cable after diagnostic operation. Use of
defective probe may cause electric shock.
3. Do not knock or bump the probe, or it may be damaged and cause electric shock.
4. Unauthorized disassembly of the probe shall be prohibited as it may cause electric shock hazard.
Attention:
1. Turn off the ultrasonic system before disconnecting the probe. Disconnecting the probe
with system power on may damage the system or probe.
2. Before disconnecting the probe, place the cable and probe on a stable and leveled position
so that the probe may not be damaged or injury person by unexpected fall.
3. Freeze the instrument when instrument is start-up without operation to increase of service life of
probe.
4. Repeat available machine time should be more than 5 minutes to avoid turn on/off power
supply in short time.