115V version
•
Voltage input 115V~ 60Hz according to your power supply
•
Power 1000VA/600W (minimum)
•
Capacity 15 minutes @ 500W (minimum)
•
Output 115V~±10% 60±1Hz
•
Transfer time 6ms (max)
Connect the UPS supplier socket (connector type IEC320–C13) to the UPS power input (yellow A).
Maximum current 4A (230V~), frequency same as local.
Connect the UPS power output to the UPS side supply (connector type IEC320-C14) (blue B).
Finally connect the socket to main power supply connector input (3): now it is possible to turn DecalMATE
ON.
2.7. Remote alarm connection
DecalMATE is also provided with two auxiliary outputs in order to connect external devices:
Output
Contact
Status
1
Identifies the DecalMATE
alarm switch
1
Normally CLOSED
2
Normally OPEN
3
COMMON
2
Identifies the DecalMATE
end program switch
7
Normally OPEN
8
COMMON
For safety reasons and for preventing any damage to the internal relays, any external
devices that need to be connected to DecalMATE unit must have a maximum Voltage
lower than 40 VDC – 1A.