DecalMATE
MM112-003 - Operator Manual
31
2.14. Owner settings
This function is available only for SERVICE and ADMINISTRATOR. Press Settings from the MAIN
FUNCTIONS screen, then Owner.
The OWNER SETTINGS screen appears as shown in the picture below.
GENERAL
Select General to modify:
Institute and Department name where unit is installed.
Reagent Cassettes Alert Status: this number indicates that
the reagent status is near expiration (on the base of the
number of cassettes processed) and the reagents status icon
turns from green to yellow (see chapter 4).
Reagent Cycles Alert Status: this number indicates that the
reagent status is near expiration (on the base of the cycles
run) and the reagents status icon turns from green to yellow
(see chapter 4).
Reagent Days Alert Status: this number indicates that the
reagent status is near expiration (on the base of the days of
use) and the reagents status icon turns from green to yellow
(see chapter 4).
VENTILATIONS
Select Ventilations to manage exhaust fumes fans of cavity
(working area) and tanks (holder), choose either:
Always active: both exhaust fans are activated at the
switching ON of the unit.
Always off: both exhaust fans are always OFF, even during
the processing.
NOT SUGGESTED.
Risk
of
fumes
in
the
laboratory
environment.
Processing and cover opening: the cavity ventilation is
activated only during the processing and when opening the
covers while the tanks ventilation is always active.
NOT SUGGESTED.
Risk
of
fumes
in
the
laboratory
environment.