DecalMATE
MM112-003 - Operator Manual
55
2. Remove the EXPIRED TANK from the position indicated by the software (number 1 in this case) and
press “Done” (see next figure) to confirm.
3. Dispose of the exhausted reagent
A. Unscrew the black cap
B. Remove the pipe
C. Pour the exhaust reagent for disposal
A
B
C
4. Load the tank with the FRESH REAGENT (Formalin in this case).
A. Put the tank in the filling position (horizontal)
B. Introduce the funnel
C. Fill the tank making sure the level REACHES THE MAXIMUM (MAX) MARK (filling label)
D. Reinsert the pipe inside the tank
E. Screw the black cap
F. Put the tank in the working position (vertical)
A
B
C
D
E
F
5. Put the tank containing fresh reagent into the position indicated by the software (number 1 in this
case). Select the reagent in the drop-down menu. The software, as a default, suggests filling with the
same reagent already loaded into the tank. To load a different reagent, select it from the drop-down
menu (figure below).
Press “Done” to confirm and return directly to the REAGENT CONFIGURATION screen.