DecalMATE
MM112-003 - Operator Manual
33
3. OPERATE WITH DecalMATE
The DecalMATE unit has dedicated software, controlled by the LCD touchscreen terminal.
Heating power is auto-tuned by a PID controller, which allows the operator to set up the desired real
parameters (Time and Temperature) by simply drawing a curve.
The unit will always reproduce the same conditions of temperature in the set times, by automatically
adjusting the power level.
The following sections show how to run and modify the program values with the DecalMATE software. Use
the touch screen to select the desired menu.
3.1. Loading cassettes into the rack
Load the RACK with cassettes and position the CASSETTE FIXING DISC on top.
3.2. Run a program
At the switch-on the DecalMATE MAIN SCREEN or FAVOURITE SCREEN appears (see chapter 2.9).
The fixing disc must always be used, in order to avoid possible spilling out of cassettes
from the rack during processing.
FIXING
DISC