DecalMATE
MM112-003 - Operator Manual
58
The screen reports that a power failure occurred, with following details:
•
Date and Time: when the program stopped
•
Processing Cycle: Run ID, protocol name and size
•
Interrupted phase
•
Interrupted after
The operator is then able to Unload Reagent, Unload cassettes and then CONTINUE the processing by
STEP START (for further details see chapter 3.2.4), according to the indications of the last step performed
by the unit prior to power failure.
AFTER THE 3 MINUTES WITHOUT POWER, WITH UPS
The system automatically puts the specimens rack in “Safe mode”, according to the following schedule:
PHASE
CONDITION
POWER
FAILURE
PROCEDURE
PHASE1
POWER FAILURE
PROCEDURE
PHASE2
POWER FAILURE
PROCEDURE
PHASE3
•
Fixation
•
Water rinsing 1-
2
LOADING
STOP THE
PHASE
UNLOAD THE
SOLUTION
RELOAD SAME
REAGENT
•
Fixation
•
Water rinsing 1-
2
PHASE
RUNNING
STOP THE
PHASE
N.D.
N.D.
•
Fixation
•
Water rinsing 1-
2
UNLOADING
THE
SOLUTION
STOP THE
PHASE
UNLOAD THE
SOLUTION
RELOAD SAME
REAGENT
•
Decalcification
ALL PHASES
STOP THE
PHASE
UNLOAD THE
SOLUTION
IF WATER RINSING
2 EXISTS, IT IS
RELOADED
Or
IF WATER RINSING
1 ONLY EXISTS, IT
IS RELOADED
Or
IF WATER RINSING
1 AND 2 DO NOT
EXIST, IT RELOAD
SAME REAGENT