B.1. HOW TO MANAGE AN ALARM
CODE
DESCRIPTION
WHAT TO DO
1
Serial communication fault
1. Press 'Clear' to reset the alarm.
2. Reboot the system.
3. Call your customer service if the issue can’t be fixed and press
'Help Desk' to manage specimens while the Alarm is
occurring.
2
Watch-dog fault
1. Press 'Clear' to reset the alarm.
2. Reboot the system.
3. Call your customer service if the issue can’t be fixed and press
'Help Desk' to manage specimens while the Alarm is
occurring.
12
Load reagent time-out
1. Press 'Silent' to hide the alarm.
2. Check that the volume of reagent in the tank is enough to
reach the level sensor. Then press 'Clear' to repeat the loading
procedure.
3. Check that the reagent tanks are rightly inserted (hoses
correctly inserted in the tanks). Then press 'Clear' to repeat
the loading procedure.
4. Check that the reagent tanks are well cleaned from any
deposits on the bottom of them. Then press 'Clear' to repeat
the loading procedure.
5. Check that the reagent tank hoses are not bent or occluded by
anything. Then press 'Clear' to repeat the loading procedure.
6. Check that the hose connector is clean and properly inserted.
Then press 'Clear' to repeat the loading procedure.
7. Call your customer service if the issue can’t be fixed and press
'Help Desk' to manage specimens while the Alarm is
occurring.
13
Unload reagent time-out
1. Press 'Silent' to hide the alarm.
2. Check that the bottom of the cavity (drainage hole) is not
occluded by anything. Then press 'Clear' to repeat the
unloading procedure.
3. Check that the reagent tanks hoses are not bent or occluded
by anything. Then press 'Clear' to repeat the unloading
procedure.
4. Check that the hose connector inside the tank is clean and
properly inserted. Then press 'Clear' to repeat the unloading
procedure.
5. Call your customer service if the issue can’t be fixed and press
'Help Desk' to manage specimens while the Alarm is
occurring.
15
Safety shield open
1. Press 'Silent' to hide the alarm.
2. Check that the safety shield is properly closed. Then press
'Clear'.
3. Call your customer service if the issue can’t be fixed and press
'Help Desk' to manage specimens while the Alarm is
occurring.
16
Tanks door open
1. Press 'Silent' to hide the alarm.
2. Check that the tanks door is properly closed. Then press
'Clear'.
3. Call your customer service if the issue can’t be fixed and press
'Help Desk' to manage specimens while the Alarm is
occurring.