DecalMATE
MM112-003 - Operator Manual
37
Particular attention must be paid to the level of reagent inside the tanks on completion of
each process to avoid low reagent situation when starting a new program.
At the end of the process, a beep sounds and the button "Check status" will flash.
Press the "Check status" icon to check the status of decalcification of the samples.
The following screen appears:
After checking the samples on the blue plastic area (see chapter 3.2.3) select:
“Run completed”: Decalcification completed then the process
will end and you can proceed with the discharge of the reagent
(reagent Unload).
or
“Go back to the decalcification”: the samples are not
completely decalcified and then restart the process by
downloading the "Water rinsing 2" and reloading the
decalcification (Decalcification Nr.2).
At every restart of decalcification, the name of the phase will
change as Decalcification nr.2, Water rinsing 2 Nr.2,
decalcification nr.3, Water rinsing 2 nr.3, etc.
It is possible to repeat this operation until 5 times.