The following window appears: set the COMPLETION TIME. In fact, DELAY START function allows setting
when the program has to finish, without modification of the existing protocol. The system keeps samples
immersed in formalin at ambient temperature (PRESOAKING TIME) and automatically calculates the starting
time to complete the processing as set by USER.
During the PRE-soaking you can add cassettes by clicking the "Add extra cassettes" button:
Particular attention must be paid to the level of reagent inside the tanks on completion of
each process to avoid low reagent situation when starting a new program.
3.3. Retrieving saved processes
Select Log Book button from the MAIN FUNCTIONS screen, then select Autosaved Programs icon.
The following screen appears showing on the right all the SAVED PROCESSING FILES, sorted by DATE
(on the left). The dates are automatically created by the software.