DecalMATE
MM112-003 - Operator Manual
18
2.5. Power supply
Power supply line must have a ground connection (yellow/green, green or bare wire).
Do not use the blue/white wire (neutral of power line). Power supply neutral has to be referred to ground.
Carefully check that:
•
A strong ground connection is made available in the lab to assure the unit works correctly. Do not
connect a Milestone unit without ground connection.
•
A dedicated ground connection direct from the main electric box (cabinet) of the laboratory to our unit
is ideal to eliminate any possible interference/electronic noise originated by other instruments
supplied with the same line.
•
Power supply line wires (their size and distance from the switch board) used in the laboratory are
compliant with the label positioned on the back side of our units.
•
Plug/socket is visible during the normal use and easily accessible.
Power line protection:
115V: Miniature circuit breaker curve C, 16A, interruption more than 6000A.
230V: Miniature circuit breaker curve C, 10A, interruption more than 6000A.
Residual current device: 30mA (Class I device).
230V version: a schuko power cord is provided.
The 115V a NEMA 5-15P power cord is provided.
Before connecting the plug to the laboratory’s socket CHECK that the power supply
voltage of the unit matches the power supply line of the laboratory.
Damages occur when wrong power supplies the instrument.
2.6. Electrical installation
2.6.1. Back side connection plate
Description:
1. Ethernet/LAN plug
2. Remote/external alarm connector
3. Main power supply connector input 230V – 115V
4. UPS side supply (controller power supply input)
5. UPS supplier socket (energy output to local UPS unit)
6. Speaker
4
3
5
2
1
6