DecalMATE
MM112-003 - Operator Manual
20
Connect the by-pass cable (CODE 66643) from (5) UPS supplier socket to (4) UPS side supply.
Finally connect the socket to main power supply connector input (3): now it is possible to turn DecalMATE
ON.
Do not connect UPS supplier socket (4) directly to the laboratory socket power supply:
this input power supply complies to CISPR 11 class A and may cause electromagnetic
interference to other devices.
2.6.4. Power supply connection (Local UPS mode)
Milestone can supply only UPS at 230V~ 50/60Hz (CODE 66100) with the following technical specifications:
230V version (CODE 66100)
•
Voltage input 230V~ 50/60Hz according to your power supply
•
Power 1000VA/600W (minimum)
•
Capacity 15 minutes @ 500W (minimum)
•
Output 230V~±10% 50/60±1Hz
•
Transfer time 6ms (max)
•
With cord set for connections
•
Space 405
×
205x145mm
•
Weight 10kg (22lb)
UPS unit at 115V~ 60Hz must be provided locally with the following minimum technical specifications:
Characteristic: Milestone s.r.l. strongly recommends this type of connection as it protects DecalMATE
from power supply failure. In this case, in fact, DecalMATE cannot process, but can execute all others
procedures. In fact “UPS side supply” provides energy at logic circuits.
Even when DecalMATE is directly plugged into the main hospital UPS line or power emergency line, we
observed occurring of delays or discontinuity of a few seconds/minutes of the power supply. In this case
DecalMATE stops working and the standard power failure procedure, that puts tissues in safety, is not
activated. For this reason, we highly recommend to install a UPS module on any DecalMATE unit.
4
3
5
2
1
6
Summary of Contents for DecalMATE
Page 1: ...DecalMATE Automatic Fixation Decalcification System OperatorManual MM112...
Page 2: ......
Page 50: ...DecalMATE MM112 003 Operator Manual 50 1 2 3 4 5 6 7 8 9 10...
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......