DecalMATE
MM112-003 - Operator Manual
67
7.2.1. Rack cleaning procedure
Clean the rack with paper and detergent, to remove any tissue and reagent residues.
7.3. Daily
It’s required to check the presence of any residual reagent in the drawer. In this case, empty it: take it out
and pour the residual reagent in a dedicated container.
7.4. Twice a week
Clean the area below the tanks and the related pipes.
7.5. Weekly
Clean the magnets behind the reagents doors and the magnet near under the transparent Perspex cover.
Avoid the use of acids or solvents.
7.6. Yearly
Contact your SERVICE support for preventive maintenance contract.