Appendix A
AccuPyc II 1340
A-8
Oct 2013
MAN_ERR:
Pressure overrange
No A.C. Found
No collected data to report,
or all cycles excluded
Printer port
not responding
Queuing print job
[ESCAPE] to cancel
Cause:
The fill valve was left open until the maximum system pressure was
exceeded.
Action:
Close the fill valve and open the vent and expansion valves. Allow the
pressure to stabilize.
Cause:
The control module cannot detect an analysis module.
Action:
Contact your Micromeritics service representative.
Cause:
You requested a report in which there is either no data available or data
have been excluded via review mode.
Action:
Initiate an automatic operation or return to review mode and include at
least one cycle.
Cause:
Your printer cable is not connected or the printer is not turned on.
Action:
Check to make sure the printer is properly connected to the
pycnometer, is turned on, and is on line.
Cause:
Status message displayed when a report is being printed.
Action:
None; this is a status message only.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...