AccuPyc II 1340
Appendix J
June 2013
J-3
10-, 3.5-, and 35-cm
3
Inserts
Simply remove the chamber cap and place the insert into the sample chamber. The insert
should fit snugly in the chamber. Place the appropriate sample cup into the well of the insert.
Be sure to install the fritted filter cap (if used) before closing the sample chamber. A fritted fil-
ter cap is used to constrict gas flow, and is included with some inserts. The fritted filter cap for
the 10-cm
3
insert fits on the top of the sample cup. The fritted filter caps for the 1- and 3.5-cm
3
inserts fit on the top of the insert. Fritted filter caps are not available for the 0.1- and 35-cm
3
inserts.
Do not force the insert or cup into the openings; the instrument, insert, or cup may be
damaged.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...