AccuPyc II 1340
System Commands
Aug 2013
3-27
Temperature
This operation typically is performed by a Micromeritics service representative.
Pressure
This calibration requires a specially designed chamber cap equipped with a pressure gauge and can
only be performed by Micromeritics service personnel.
Place the calibration standard(s) in the cup in the sample cell
chamber and replace the chamber cap; then press
ENTER
to
calibrate the temperature offset (
Cal2
).
Calibration continues and operational status messages are
continually displayed. After the operation is complete, the
pycnometer returns to the
Reload
prompt.
For best results, sample volume should be approximately
equal to the volume of the calibration standard. Therefore, if
you are calibrating the 10-cm
3
pycnometer for use with
smaller sample volumes, you may wish to use only one stan-
dard. If you use one standard, you should enter half of the
value at the
Volume of cal std
prompt (shown on the previ-
ous page).
Press
Alt + CLEAR
if you wish to cancel the operation.
Enter the temperature obtained from a reference temperature
sensor.
Press
ENTER
; you automatically return to the
Reload
prompt.
Press
ALT + CLEAR
if you wish to cancel the operation.
Enter the pressure value obtained from the external pressure
device, then press
ENTER
; you automatically return to the
Reload
prompt.
Press
Alt + CLEAR
if you wish to cancel the operation.
Unit[n]
SN1234 10 cm3
X - X - X
Insert Cal Std
[ENTER] to start
Unit[n]
SN1234 10 cm3
Calibrate Temperature
Temperature:
(user-entered)
Unit[n]
SN1234 10 cm3
Calibrate Pressure
Pressure:
(user-entered)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...