Troubleshooting
AccuPyc II 1340
5-2
Aug 2013
What Happened
Why
What To Do
Specified pressure not
reached or maintained.
(continued)
There are gas leaks in
the cell chamber or
expansion chamber.
Check the pycnometer for leaks (refer to
Checking the Cell and Expansion
Chambers for Leaks
).
The Helium tank is
empty.
Replace the tank. Minimum
recommended pressure is 200 psig.
The cylinder Shut-Off
valve is closed.
Open the Shut-Off valve.
The pressure regulator is
defective, or the pressure
is set too low.
Ensure that the regulator is set properly.
Call a Micromeritics Service Represen-
tative, if necessary.
The zero offset (of
pressure transducer) is
too low.
Check offset by opening the chamber
cap. If the offset is negative, run a new
zero offset (refer to
Dust filter on 2000 cm
3
unit is clogged or
leaking to atmosphere.
Clean the dust filter. Refer to
the Dust Filter (2000-cm3 units only)
.
Helium drained from
tank.
Leaks in the gas line
connection.
Pressurize the system. Close, then open
the Cylinder Shut-Off Valve. If the
needle on the pressure gauge on the gas
cylinder jumps abruptly, a leak in the
gas line connections may be indicated.
Check gas line connections.
The pycnometer was left
in manual mode with all
valves open, or the fill
valve is open and the
chamber cap is off.
Close all valves, then attach a new tank
of helium.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...