6. Press
CHOICE
until the average measured volume is displayed; record the value.
7. Press
Alt + 2
to access
Setup
.
8. Press
CHOICE
until
Calibration Data
is displayed.
9. Press
ENTER
; the
Chamber Insert
prompt is displayed.
10. Ensure that
None
is selected, then press
ENTER
to display the
Cell Volume
prompt.
11. Subtract the average volume of the 0.1-cm
3
insert/cup (recorded in Step 6) from the cell
volume displayed in the prompt; record this value.
12. Press
ENTER
to display the
Expansion Volume
prompt; record the value displayed.
13. Record the value displayed for the expansion volume, then press
ENTER
. This returns you
to the
Calibration Data
prompt.
14. Press
ENTER
; the
Chamber Insert
prompt is displayed.
15. Press
CHOICE
until
0.1 cm
3
is displayed, then press
Enter
to display the
Cell Volume
prompt. Enter the value you recorded in Step 11.
16. Press
ENTER
to display the
Expansion Volume
prompt; enter the value you recorded in
Step 12.
17. Press
SAVE
to save the information you entered and return to the
Reload
prompt.
The pycnometer is now ready for analyses using the 0.1-cm
3
insert/cup.
Unit[n]
SN1234 10 cm3
Set-up Type?
Calibration Data
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...