AccuPyc II 1340
Reviewing Data
June 2013
4-7
Reviewing Data
The
Review
function enables you to review and edit the results of the last operation – analysis
or calibration – along with its entered parameters. It is important that you review and print the
report (if desired) before starting another operation. Data are no longer available for review on
the display when another operation is started; however, data results are saved in the control
module and can be reviewed using a web browser (refer to
).
Analysis Data
The prompts that display during the review for an analysis depend on the options you choose
in
Set Up > Analysis Parameters
and
Report Options
. Therefore, all the prompts given in
this procedure may not apply to your analysis.
Values in the fields of these prompts may be edited if desired.
1.
Press
Alt + 5
; the
Sample ID
prompt containing the sample identification is displayed.
2.
Press
ENTER
; the
Description
prompt containing an additional description is displayed.
This prompt typically is not used unless you have a keyboard attached to your analyzer.
3.
Press
ENTER
; the
Sample Mass
prompt containing your sample’s mass is displayed.
4.
Press
ENTER
; the
Start
prompt containing the start date and time of your analysis is
displayed.
5.
Press
ENTER
; the
End
prompt containing the ending date and time of your analysis is
displayed.
6.
Press
ENTER
; the first of the prompts containing data is displayed.
or
The
[n]
displayed in the example prompts represents the cycle number.
7.
Continuing pressing
ENTER
to view data.
Press
CHOICE
to exclude data from report calculations; an asterisk will display
indicating that it will be excluded. Press
CHOICE
again to remove the asterisk.
8.
Press
SAVE
to return to the
Reload
prompt.
9.
For a record of analysis results, press
Alt + 6
to print to an attached printer.
Unit[n]
SN1234 10 cm3
X - X - X
Dn [n] = (density)
Dv [n] = (deviation)
Unit[n]
SN1234 10 cm3
X - X - X
V [n] = (volume)
Dv [n] = (deviation)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...