Reviewing Data
AccuPyc II 1340
4-8
June 2013
Calibration Data
The prompts that display during a review of calibration data depend on the options you specify
in
Set Up > Analysis Parameters
and
Report Options.
You may edit the information in the prompts if desired.
1.
Press
Alt + 5
; the
Chamber Insert
prompt showing the insert (if used) is displayed.
2.
Press
ENTER
; the volume of the calibration standard is displayed.
3.
Press
ENTER
; the
Which Chamber
prompt is displayed.
4.
Press
CHOICE
to select the chamber for which you wish to view volume results.
5.
Press
ENTER
; a prompt displaying the starting time and date is displayed.
6.
Press
ENTER
; a prompt displaying the ending time and date is displayed.
7.
Press
ENTER
; the first of the prompts containing data is displayed.
or
The [n] displayed in the example prompts represent the cycle number.
8.
Continuing pressing
ENTER
to view all of the data.
Press
CHOICE
to exclude data from report calculations; an asterisk will display
indicating that it will be excluded.
9.
Press
SAVE
to return to the
Reload
prompt
10. For a record of calibration data, press
Alt + 6
to print to an attached printer.
Unit[n]
SN1234 10 cm3
X - X - X
Cell [n] = (volume)
Dev [n] = (deviation)
Unit[n]
SN1234 10 cm3
X - X - X
Exp [n] = (volume)
Dev [n] = (deviation)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...