10. Press
7
to close the Fill valve;
X - O - X
.
11. Observe the pressure display. After an equilibration period (about 20 to 30 seconds), the
pressure should not vary more than 0.005 psig/min.
If the pressure varies more than 0.005 psig/min, temperature instability or a leak may be
indicated. Vent the system, then repeat steps 8 through 11 several times to verify that a
leak is indicated. If a leak is indicated, call a Micromeritics Service Representative.
Cleaning the Pycnometer
The exterior casing of the pycnometer may be cleaned using a clean cloth, dampened with
isopropyl alcohol (IPA), a mild detergent solution, or a 3% hydrogen peroxide solution.
Do not immerse the pycnometer or the power cord in any liquids. Doing so could
result in electrical shock to personnel or damage to the unit.
Do not allow liquid to penetrate the casing of the pycnometer. Doing so could result in
damage to the unit.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...