Appendix A
AccuPyc II 1340
A-4
Oct 2013
CAL1ERR: Eq failure
NN/ZZ cycles completed
CAL1ERR: Fill failure
NN/ZZ cycles completed
CAL1ERR: Overrange
NN/ZZ cycles completed
Cause:
The empty cell chamber failed to equilibrate in 1000 seconds during
the first pass of calibration.
Action:
Check the system for leaks (refer to
Checking the Cell and
Expansion Chambers for Leaks
in Chapter 5).
Cause:
There was insufficient pressure to allow filling within five minutes
during the first pass of calibration.
Action:
Open the tank valves if shut, increase the regulator pressure or, if there
is insufficient helium in the tank, obtain a new tank of helium.
(Minimum recommended tank pressure is 200 psig.)
Cause:
A pressure overrange occurred during the first pass of calibration
because the regulator pressure is set too high.
Action:
Adjust the pressure so that it is equal to the higher of the two fill
pressures specified in the analysis parameters plus 2.0 psig.
Cause:
A pressure overrange occurred during the first pass of calibration
because an error occurred in the pressure measurement electronics.
Action:
Call a Micromeritics Service Representative.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...