AccuPyc II 1340
System Commands
Aug 2013
3-29
Reset Pressure Calibration
This option can be used when you suspect the pressure is not being reported correctly by the analyzer.
For example: a zero, negative, or unreasonable high reading is being consistently reported. Resetting to
nominal may restore normal pressure readings but a proper calibration should be performed.
Enter the serial number, then press
ENTER
.
These prompts appear in succession.
Remove the USB stick; the system automatically returns to
the
Reload
prompt.
Press
CHOICE
until the desired option is displayed:
Press
ENTER
.
This resets the pressure calibration to nominal values;
temperature and volume calibrations are not affected.
Nominal values may be sufficient for your application.
However, you should schedule a proper pressure calibration
as soon as possible. Press
CHOICE
.
Choices:
Yes, No
Select
Yes
to reset the calibration to nominal values. Contact
your service representative and schedule a proper pressure
calibration since data accuracy may be compromised.
Select
No
if you wish to contact your service representative
to schedule a proper pressure calibration.
Press
ENTER
.
Unit[n]
SN1234 10 cm3
X - X - X
Load cal. for SN:
Unit[n]
SN1234 10 cm3
X - X - X
Calibration for
SN (nnnn) loaded
Unit[n]
SN1234 10 cm3
X - X - X
Remove USB stick
Unit[n]
SN1234 10 cm3
X - X - X
Calibration Type?
Reset Pressure Cal.
Unit[n]
SN1234 10 cm3
X - X - X
Reset Pressure Cal.?
No
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...