AccuPyc II 1340
Appendix A
Oct 2013
A-1
A. ERROR MESSAGES
This appendix contains an alphabetical listing of the messages that may appear in the display
of the pycnometer.
Operational and report error messages generated during an analysis or calibration are placed in
a queue, along with report data. Error messages are indicated with an asterisk at the
Reload
prompt after the operation is completed.
Messages and data are placed in the following order:
• Operational error messages
• Report error messages
• Report data
Press
CHOICE
to cycle through the messages and to the data. If you wish to delete the message
after viewing its contents, press
CLEAR
. If you aren’t clear as to why the message was
displayed, look it up in this appendix for a cause and a resolution before deleting. Messages
that are not cleared will remain in the queue until the next analysis or calibration is performed;
at that time, any remaining messages will be cleared automatically.
When you see NN/ZZ in a message listed below,
NN
= number of cycles completed and
ZZ
= number of cycles requested.
Add Unit
Cause A:
A new analysis module has been attached to the instrument where
previously a different module had been attached.
Action A:
Press
Choice
and select
Yes
to add the new analysis module.
Cause B:
An incorrect serial number was read by the analysis module.
Action B:
Press
Choice
and select
No
to use the serial number and calibration
information stored in the command module. If this message displays
every time you turn on the instrument, contact your service
representative.
Unit[n]
SN1234
10 cm3
X - O - X
Reload *
P = (current pressure)
An asterisk (*) indicates error message
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...