AccuPyc II 1340
Appendix B
Aug 2013
B-1
B. CALCULATIONS
This appendix contains the calculations used in reports available in the AccuPyc program.
Calibration
Volume calibration uses the ultraprecise method of separate adjustments for the offset and
scale factor.
Volume Offset
–V
sampempty
is reported as the offset in volume calibration reports.
Volume scale
is reported as the scale factor in volume calibration reports.
where
V
cel
= sample chamber volume
V
celprev
= previously stored cell volume
V
sampempty
= average
V
samp
from volume offset calibration analysis (no calibration ball)
V
calib
= calibration ball volume
V
sampball
= average
V
samp
from volume scale calibration analysis (with calibration ball)
V
exp
= expansion chamber volume
V
expprev
= previously stored expansion volume
V
cel
V
celprev
V
sampempty
–
=
V
cel
V
celprev
V
calib
V
sampball
-----------------
=
V
exp
V
expprev
V
calib
V
sampball
-----------------
=
V
calib
V
sampball
-------------------
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...