AccuPyc II 1340
Maintenance
Aug 2013
5-5
Replacing the Chamber Cap O-Ring
Fine fibers and particles between the O-ring and its sealing surfaces can cause leaks, as can
scratches or cuts in the O-ring or in the metal surfaces. When it is necessary to replace the
O-ring, follow these steps.
1.
Using a pointed tool, carefully remove the O-ring from its groove in the cap. A small
niche is provided at the groove for placement of the tool.
2.
Clean the groove in the chamber cap using a small brush or lint-free tissue moistened
with isopropyl alcohol.
3.
Allow the chamber cap to dry thoroughly.
4.
Place a small amount of Dow Corning High vacuum grease (or equivalent) on your index
finger. Touch your thumb with your index finger to place grease on both fingers.
Be careful not to scratch the metal surface of the chamber cap. Scratches could result
in an imperfect seal.
Apply the grease sparingly; too much grease may alter cell volume, while too little
grease results in an imperfect seal and leaks.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...