The AccuPyc is a standard DTE device. The RS-232 port on the rear panel of the AccuPyc can
be used to attach an analytical balance for transfer of sample weight, or for transmitting data
through a serial line to a computer.
The pin assignment for the RS-232 port is shown in the following table; be sure your receiving
device is configured to interface with these assignments. Any signals that are not listed in the
table will be ignored by the AccuPyc.
The AccuPyc uses the DTR and DSR signals for hardware flow control. Be sure that your
serial device provides these signals. For example; if attaching to a PC (also a DTE device), use
a null modem cable which includes the signals designated in the above table.
If you experience a problem with transmission, ensure that the signals are set up properly. If
the signals are correctly configured, contact the receiving device manufacturer for assistance.
Pin
Signal
Description
Data Direction
2
RXD
Receive data
Into AccuPyc
3
TXD
Transmit data
From AccuPyc
4
DTR
Data terminal ready
From AccuPyc
5
GND
Ground
—
6
DSR
Data set ready
Into AccuPyc
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...