Cleaning the Dust Filter (2000-cm
3
units only)
Dust from the sample may be carried into the instrument plumbing. A dust filter protects the
valves to minimize valve leaks. When it is necessary to clean the filter, follow these steps:
1.
Close the gas line (Fill) valve. Open the expansion and vent valves to ensure that the
system is not under pressure.
2.
At the lower rear of the instrument, remove the two screws which hold the dust filter
cover and remove the cover plate.
3.
Remove the spring and the filter. Remove the small O-ring inside the filter port.
4.
Clean all parts with alcohol. The filter may require sonication to dislodge dust. It should
be replaced if it becomes severely clogged.
5.
Inspect the O-rings before re-assembly. A very thin coating of vacuum grease may be
used.
6.
Reassemble.
7.
Check for leaks:
a.) Fill the system with gas.
b.)
Open expansion valve.
c.)
Check for leaks by applying a leak detecting liquid around the dust filter cover.
Filter port
Coverplate
O-ring (size 006)
Retaining screws
Filter
Spring
O-ring
(size 011)
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...