SYS_ERR: Power Fail
NN/ZZ Cycles Completed
TRN_ERR: Timeout
failed to respond
Updating Unit [n]
failed: Press [CLEAR]
USR_ERR: Cal std
10% of full-scale
Cause:
A power failure occurred and when power resumed, the automatic
operation was canceled.
Action:
Restart the automatic operation if desired.
Cause:
The receiving device took longer than 10 seconds to acknowledge
receipt of data from the pycnometer.
Action:
Make sure the receiving device is properly connected to the
pycnometer RS-232 port and is turned on. Verify that the serial I/O
parameters controlling the receiving device correspond with the data
transmission parameters in the set up mode.
Cause:
One of the analysis modules has malfunctioned.
Action:
Press
CLEAR
and follow the instructions displayed in the keypad
window.
After disconnecting the faulty unit (as instructed), you may reconnect
the control module with the remaining analysis module(s) and continue
using your system.
Cause:
You tried to enter a calibration standard volume that is less than 10% of
the nominal full-scale volume.
Action:
Enter a volume that represents at least 10% of the nominal cell
chamber volume.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...