CAL2ERR: Underrange
NN/ZZ cycles completed
DATA_ERR: Chamber
volumes must be > 0
DATA_ERR:
No data to compute
Cause:
A pressure underrange occurred during the second pass of calibration
because an error occurred in the pressure measurement electronics.
Action:
Call a Micromeritics Service Representative.
Cause:
According to the calibration data, the cell volume or expansion volume
for the selected insert is zero or a negative value.
Action:
Perform the following steps:
1.
Select
Set Up > Calibration Data
.
2.
At the
Cell volume
prompt, enter the volume of the sample cell
(or insert).
3.
At the
Expansion volume
prompt, enter half of the sample cell
volume.
4.
Perform a volume calibration.
Cause:
An automatic operation was canceled before a complete cycle could be
completed.
Action:
Restart the automatic operation.
Cause:
All the cycles have been excluded in review mode.
Action:
Return to review mode and include some cycles.
Summary of Contents for AccuPyc 2 1340
Page 1: ...AccuPyc II 1340 Operator s Manual V1 09 134 42801 01 Oct 2013...
Page 2: ...Micromeritics Instrument Corporation 2006 2013 All rights reserved Printed in the U S A...
Page 4: ......
Page 8: ...Table of Contents AccuPyc II 1340 iv Aug 2013...
Page 18: ...Specifications AccuPyc II 1340 1 10 Aug 2013...
Page 72: ...System Commands AccuPyc II 1340 3 38 Aug 2013...
Page 84: ...Reviewing Data AccuPyc II 1340 4 12 June 2013...
Page 100: ...Ordering Information AccuPyc II 1340 6 6 Sep 2013...
Page 116: ...Appendix B AccuPyc II 1340 B 4 Aug 2013...
Page 120: ...Appendix C AccuPyc II 1340 C 4 June 2013...
Page 128: ...Appendix D AccuPyc II 1340 D 8 Nov 06...
Page 136: ...Appendix G AccuPyc II 1340 G 2 Oct 08...
Page 138: ...Appendix H AccuPyc II 1340 H 2 Nov 06...
Page 152: ...Appendix K AccuPyc II 1340 K 4 Apr 07...
Page 157: ...AccuPyc II 1340 Index June 2013 Index 5 Zero pressure offset 1 5 Zero pressure transducer 3 37...